[ Identification | Description | Examples | Input parameters | Links ]

The SOLEIL_SWING Instrument

A simple model for the SWING beam-line at SOLEIL (small angle scattering, SAXS).

Identification

Description

The SWING beamline targets soft condensed matter, conformation of
macro-molecules in solution (BioSAXS) and material sciences. Our experimental
set up allows simultaneous small-angle X-ray scattering (SAXS) and wide-angle
X-ray scattering measurements (WAXS) in the 5-16 keV energy range. Anomalous
scattering experiments can also be performed. A very large variety of types of
samples can be studied, e.g., solutions, gels, amorphous solids, crystallized
solids, thanks to the diversity of the proposed sample environments.

This model may use any PDB file as sample, or an ideal sphere colloid for testing.

Position | Element
---------|--------------------------------------------------------------------
0        | U20 undulator
11.7     | Diaphragm providing HxV=1x0.5 mm beam
20       | Si(111) DCM, 40x40x10 mm^3 SWING_E0=5-16 keV
22.5     | Focusing KB (HFM + VFM)
31       | Linear H-CRL (f=81 cm): 31 m
32       | Sample location
32.5-39  | Detector 162.5 x 155.2 mm

Examples

(Test cases in bold)

Input parameters

Parameters in boldface are required; the others are optional.
NameUnitDescriptionDefault
SWING_E0keVNominal energy at the Wiggler.13
SWING_dEkeVEnergy half-bandwidth at the Wiggler0.1
SWING_dcm_thetadegRotation angle of the DCM. 0=set from energy SWING_E00
SWING_hfm_radiusmHorizontally focusing mirror radius.495
SWING_vfm_radiusmVertically focusing mirror radius.859
SWING_mirror_grazing_angledegTilt angle of the mirrors.4e-3
SWING_samplestrSample given as a PDB file, or NULL for a 100A dilute Sphere model."6lyz.pdb"
SWING_sample_detmSample to detector distance in m.2
splits10

Links


[ Identification | Description | Examples | Input parameters | Links ]

Generated for mcxtrace 3.10.0