ComponentModel of a laboratory x-ray tube, generating x-rays by bombarding a target by electrons. Given a input energy E0 of the electron beam, x-rays are emitted from the accessible emission lines The geometry of the tube is assumed to be: # The electron beam hits a slab of surface material surface at a right angle illuminating an area of width by height, # where width is measured along the component X-axis. # The centre of the electron beam at the anode surface is the origin of the component. # The Z-axis of the component points at the centre of the exit window (focus_xw by focus yh) placed at a distance dist from the origin. # The angle between the Z-axis and the anode surface is the take_off angle. For a detailed sketch of the geometry see the componnent manual. The Bremsstrahlung emitted is modelled using the model of Kramer (1923) as restated in International Tables of Crystallography C 4.1 Characteristic radiation is modelled by Lorentzian (default) or Gaussian energy profiles with line-energies from Bearden (1967), widths from Krause (1979) and intensity from Honkimäki (1990) and x-ray data booklet. Absoprtion of emitted x-rays while travelling through the target anode is included. Example: Source_lab(material_datafile="Cu.txt",Emin=1, E0=80)
Name | Unit | Description | Default | |
material_datafile | Name of datafile which describes the target material. | "Cu.txt" | ||
width | m | Width of electron beam impinging on the anode. | 1e-3 | |
height | m | Height of electron beam impinging on the anode. | 1e-3 | |
thickness | m | Thickness of the anode material slab. | 100e-6 | |
E0 | kV | Acceleration voltage of xray tube. | 20 | |
Emax | keV | Maximum energy to sample. Default (Emax=0) is to set it to E0. | 0 | |
Emin | keV | Minimum energy to sample. | 1 | |
focus_xw | m | Width of exit window. | 5e-3 | |
focus_yh | m | Height of exit window. | 5e-3 | |
take_off | deg | Take off angle of beam centre. | 6 | |
dist | m | Distance between centre of illuminated target and exit window. | 1 | |
tube_current | A | Electron beam current. | 1e-3 | |
frac | Fraction of statistic to use for Bremsstrahlung. | 0.1 | ||
lorentzian | If nonzero Lorentzian (more correct) line profiles are used. | 1 | ||
xwidth | m | Width of the anode material slab. | 0 | |
yheight | m | Height of the anode material slab. | 0 | |
exit_window_refpt | m | If set, the AT position and exit window will coincide (legacy behaviour). | 0 |
AT ( | , | , | ) RELATIVE | |||
---|---|---|---|---|---|---|